Skip Navigation



Rheumatology Advance Access published online on December 6, 2005

Rheumatology, doi:10.1093/rheumatology/kei243
This Article
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
45/5/595    most recent
kei243v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by van der Esch, M.
Right arrow Articles by Dekker, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by van der Esch, M.
Right arrow Articles by Dekker, J.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The Author 2005. Published by Oxford University Press on behalf of the British Society for Rheumatology. All rights reserved. For Permissions, please email: journals.permissions@oxfordjournals.org
Received May 23, 2005
Accepted November 4, 2005

Original Article

Reproducibility of instrumented knee joint laxity measurement in healthy subjects

M. van der Esch 1 *, M. Steultjens 2, R. W. J. G. Ostelo 3, J. Harlaar 4, and J. Dekker 2

1 Jan van Breemen Institute, Center for Rehabilitation and Rheumatology, Amsterdam, The Netherlands
2 Jan van Breemen Institute, Center for Rehabilitation and Rheumatology, Amsterdam, The Netherlands; VU University Medical Center, Department of Rehabilitation Medicine, Amsterdam, The Netherlands; Institute for Research in Extramural Medicine (EMGO Institute), VU University Medical Center, Amsterdam, The Netherlands
3 Institute for Research in Extramural Medicine (EMGO Institute), VU University Medical Center, Amsterdam, The Netherlands; Department of Allied Health Care Research, Amsterdam School of Allied Health Education, Amsterdam, The Netherlands
4 VU University Medical Center, Department of Rehabilitation Medicine, Amsterdam, The Netherlands; MOVE Research Institute for Clinical Movement Studies, Amsterdam, The Netherlands

* To whom correspondence should be addressed.
M. van der Esch, E-mail: M.van.der.Esch{at}planet.nl


   Abstract

Objective. To determine the reproducibility of frontal plane knee joint laxity measurement through the assessment of intra- and inter-rater reliability coefficients and intra- and inter-rater agreement coefficients.

Methods. Two raters independently assessed the laxity of the knee joint in the frontal plane by three repeated measurements. Fourteen days later the assessment was repeated. Complete data were obtained from 20 healthy subjects. Laxity was assessed using a device which consisted of a chair with a free-moving arm that supported the subject's lower leg. Medial and lateral loads were applied, resulting in a varus and valgus movement in the knee joint. The intra- and inter-rater reliability coefficients [intraclass correlation coefficients (ICC)] were estimated, as were the intra- and inter-rater agreement parameters [standard error of measurement (SEM) and minimal detectable difference (MDD)].

Results. Adequate intra-rater reliability (ICC>0.80) was calculated for each rater's measurements of laxity. The inter-rater reliability was less adequate (ICC=0.65) when calculated using the first day's measurements. However, inter-rater reliability was adequate (ICC=0.88) when calculated using the day 14 measurements. The intra-rater measurement error calculated across occasions was 1.3° for individual subjects. This resulted in an MDD of 3.7°. The inter-rater measurement error, i.e. the SEM and MDD, was higher (1.5° and 4.3°, respectively).

Conclusions. Intra-rater reliability of knee joint laxity measurement is good. Adequate training of raters establishes the basis for good inter-rater reliability. In clinical trials, it is preferable for one trained rater to perform the laxity measurement. The measurement of knee joint laxity is limited due to its relatively high measurement error in individual subjects; therefore, this measurement should be restricted to group assessment rather than individual patient assessment.

Keywords: Reproducibility; Reliability; Knee joint laxity; Osteoarthritis..
Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
Rheumatology (Oxford)Home page
S. F. E. van Weely, C. J. van Denderen, I. E. van der Horst-Bruinsma, M. T. Nurmohamed, B. A. C. Dijkmans, J. Dekker, and M. P. M. Steultjens
Reproducibility of performance measures of physical function based on the BASFI, in ankylosing spondylitis
Rheumatology, October 1, 2009; 48(10): 1254 - 1260.
[Abstract] [Full Text] [PDF]


Home page
Neurorehabil Neural RepairHome page
H.-M. Chen, C. C. Chen, I-P. Hsueh, S.-L. Huang, and C.-L. Hsieh
Test-Retest Reproducibility and Smallest Real Difference of 5 Hand Function Tests in Patients With Stroke
Neurorehabil Neural Repair, June 1, 2009; 23(5): 435 - 440.
[Abstract] [PDF]


Home page
Neurorehabil Neural RepairHome page
Y.-W. Hsieh, C.-H. Wang, S.-C. Wu, P.-C. Chen, C.-F. Sheu, and C.-L. Hsieh
Establishing the Minimal Clinically Important Difference of the Barthel Index in Stroke Patients
Neurorehabil Neural Repair, May 1, 2007; 21(3): 233 - 238.
[Abstract] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.